25
OxyMag_rev26
4.4
Mounting the vertical support
Support (1702496) is an optional item and can be used in
ambulances, helicopters or walls of hospital environment
facilities (emergency, post-anesthetic recovery rooms, ICU,
etc).
Below is provided the sequence to assemble the support on
the wall.
1. Install the fixed support with +12V DC power (3803835) on
the wall (room, ambulance, helicopter, etc.) using 4 screws
(3003446) item 1 of the figure to the side and 4 fixing bolts
(3003447), if required (item 2 of the figure).
Figure 8: Installation of the fixed support
2. To place the ventilator on the support, follow the sequence
below:
a.
Pull the handle bracket on the wall, just above the
fixed support;
b.
Slide Oxymag down until it clicks into place;
c.
Press the safety lock of the ventilator turning the two
eccentric buttons from the top until the red dots are
invisible;
d.
Make sure that Oxymag is fixed in placed;
e.
To remove Oxymag, perform the reverse procedure.
Figure 9: Connection of the ventilator to the fixed support
Below is provided the sequence to assemble the support on the
bench:
1.
Install the fixed support with +12V DC power (3803835) on
the bench using 2 screws (3003446) item 2 of the figure to
the side.