94
OxyMag_rev26
Item
Parameter
Specification
Tolerance
Unit
Barometric Pressure
600 - 1100
---
hPa
Relative Humidity (w/o condensation)
15 - 95
---
%
4
Storage/ Transport
Temperature
-25 - 75
---
°C
Barometric Pressure
500 - 1200
---
hPa
Relative Humidity (w/o condensation)
5 - 95
---
%
5
Life time
10
---
years
6
Time to heat or cool equipment stored at extreme temperatures
to operate at 20 ° C
30
---
minutes
Internal volume of respiratory circuit components
INTERNAL VOLUME
SILICONE RESPIRATORY TUBE 22M+22M X 1,20M
AUTOCLAVABLE
407,8 mL
Y ADULT 22MM AUTOCLAVABLE WITH THERMOMETER
18,7 mL
SILICONE RESPIRATORY TUBE 15MM X 1,20 M
AUTOCLAVABLE 15F+22F
179,5 mL
INTERMEDIARY 15M+15M FOR PEDIATRIC BREATHING
CIRCUIT AUTOCLAVABLE
3,6 mL
Y PEDIATRIC 15MM AUTOCLAVABLE WITHOUT
THERMOMETER 90
16,9 mL
AIRWAY ADAPTER
9,0 mL
ADULT FLOW SENSOR - AUTOCLAVABLE
8,4 mL
PEDIATRIC FLOW SENSOR - AUTOCLAVABLE
8,0 mL
NEONATAL FLOW SENSOR - AUTOCLAVABLE
7,7 mL
Extreme conditions
•
Do not store the ventilator in environments outside the temperature, humidity and pressure
specified in 14.3.3 Physical and Environmental Specifications. Accuracy of equipment
readings may be affected.
•
AC mains power supply with voltage values below 25% may result in switching power to the
internal battery.
•
Power above 15% of the rated value may result in equipment AC / DC power failure, but the
equipment will continue to operate normally due to switching to the internal battery.