118
OxyMag_rev26
14.4
O2 galvanic cell specification
General specification
Intended use
Measure the O
2
concentration delivered from
the equipment to the patient
Measuring range
0 to 100%
Output sign
9
–
13 mV
Response time 90%
13 s
Accuracy
± 2%
Linearity
± 2%
Recommended flow rate
0.1
–
10 lpm
Data sampling rate
7 Hz
Method for calculating the gas level reading
Simple moving average (MMS) of 64 positions
acquired every 140ms
Respiratory rate
The Respiratory Rate is shown every 3 breaths
and the average value is updated with each breath.
Effects of gas and vapor interference
Gases or Steam
Gas Level
Response to 80% NO
< 5%
Response to 7,5% Halothane
< 5%
Response to 7,5% Isoflurane
< 5%
Response to 7,5% Enflurane
< 5%
Response to 9% Sevoflurane
< 5%
Response to 20% Desflurane
< 5%
Response to 10% CO
2
< 5%