24
OxyMag_rev26
message is LOW BATTERY, provide
IMMEDIATE connection of power supply
to the mains, if DISCONNECTING the
equipment from the patient is not
possible and provide appropriate means
of ventilatory support.
•
Use the appropriate breathing circuit to
the patient.
•
When using oxygen cylinder, check if
the pressure reducing valve is set to
deliver oxygen flow with pressure
according to 14.3.2 Connecting to the
Oxygen Supply. Pressures greater than
specified may damage the equipment.
4.3
Power Connection
The equipment must be connected to a three-pin grounded
power outlet fulfilling ABNT NBR 13534 standard
–
“Electrical
installations in health care facilities
–
Safety requirements”.
The internal battery to the equipment must Always be charged
and ready for use in a failure of the mains or for use in external
operations; for such, your power supply should be connected
to the mains to charge the battery even if the equipment
remains off.
After prolonged use of the equipment, using only the energy
of the internal battery, full recharge is required preparing the
equipment for the extended use.
If the equipment remains unplugged for more than a month,
the battery should be fully recharged.
Caution
•
Do not position the equipment so that it is difficult to operate the device from disconnecting of the power
supply.
Note
•
After the power supply has been interrupted and restored while the internal battery is charged while the
equipment is operating, equipment performance will not be affected and accuracy will be maintained.
•
After a long period of interruption to the power supply, connect the equipment to the mains, switch on the
equipment and wait 30 min. Perform necessary calibrations and self test.