110
OxyMag_rev26
Pneumatic diagram
1
Filter
11
Pressure measuring point
2
Pressure regulating valve 45 psi
12
Humidifier (optional)
3
Pressure sensor 150 psi
13
Breathing circuit
4
Proportional valve 30 LPM
14
Capnograph (optional)
5
Venturi system
15
HEPA or HME filter
6
Proportional valve 200 LPM
16
Proximal flow sensor
7
One-way valve (anti-asphyxiation)
17
Expiratory valve
8
O
2
cell
18
Linear actuator
9
Overpressure valve On/Off
19
HEPA filter
10
Internal flow sensor
20
Measuring point (proximal pressure differential)
21
Solenoid valve x-1 / 6 psi
Figure 31: Pneumatic Scheme of the transport ventilator