Once all ventilation parameters are set on the ventilator, the patient breathes spontaneously defining the control times of ventilation.
1
and
2
Represent spontaneous cycles without pressure support at P. High (High Continuous Airway Pressure);
3
Represents the transition of P.High to P.Low (Low Continuous Airway Pressure) synchronously;
3
to
4
Represents the time T.Low in which airway pressure release is performed;
4
Represents transition from P.Low to P. High synchronously.
Transitions of levels from P. High P.Low or P.Low P. High occur in the final quarter of T. High and T.Low, respectively, by
synchronizing with the patient’s effort. The information of what type of trigger activated the inspiratory cycle is reported
in the
message area and screen status.
•
Default values are only for initial reference. Reset the ventilation parameters as needed by
the patient.
•
To obtain DUALPAP mode with backup ventilation, select backup ventilation in the backup
parameter.
Note
•
Respiratory rate monitored is result of spontaneous breathing of the patient
•
Pressure support (ΔPS) is a value above P.High or P.Low and can be adjusted between
PMax-
P.High + 5 cmH2O e PMAX - P.High.
•
Changes in pressure levels are synchronized.