82
OxyMag_rev26
Caution
•
Not to be used for cleaning or disinfecting the phenol (> 5%) ketone, formaldehyde, hypochlorite,
chlorinated hydrocarbons, aromatic hydrocarbons, inorganic acids, and quaternary ammonium
compounds.
•
Never use saline solutions, especially sodium hypochlorite (bleach) and saline, disinfectants, hydrogen
peroxide for cleaning or rinsing the accessories.
10.1.2.3
Drying
Drying of the external parts should be done with a clean, soft and dry cloth and the drying of the internal parts should be done
so that the solution drains by gravity.
10.2
Disinfection
External Parts
The external part should be disinfected using a clean cloth moistened with alcohol 70 º.
Respiratory circuit, exhalation valve, proximal flow sensor and silicone line
After cleaning, the items should be disinfected with alcohol 70º. The external part should be disinfected using a clean cloth
moistened with alcohol 70 º and internal part by immersion. Important: Do not soak the items to be disinfected with alcohol as it can
damage the material.
After disinfection, the external parts should be dried with a clean, soft and dry cloth and the internal parts should be dried so
that the solution drains by gravity.
10.3
Sterilization
•
The components that get in touch with the respiratory gases shall be disassembled in components, as accessory
list, for cleaning and sterilization;
•
Do not use abrasive agents to carry out cleaning;
•
Do not use alcohol to clean the plastic parts;
•
Do not immerse the Oxymag in any liquid;
•
When sending the Oxymag for
maintenance or repair strictly
observe the disinfection process.
•
Equipment visibly infected by
patients fluid will be returned
without performing maintenance or
repair service.
•
This equipment and the parties must
go through a cleaning process every
time it is even used in the first use.