42
OxyMag_rev26
Attribute
Specification
Warm-up
Information on the concentration is analyzed and sent every 10
seconds.
Overall accuracy in measurements: 1 minute.
Rise time (at 10 l/min)
CO
2
≤ 90ms.
Overall System Response Time
< 1s.
Note
•
CO
2
monitoring is achieved accurately 1 minute after initialization.
Table 16: Accuracy specifications of the Capnography
Precision / Accuracy of measurements (under standard conditions):
Gas Type
Gas Type
Gas Type
CO
2
0 - 15
15 - 25
±(0.2 vol% + 2% reading)
Not specified
Note:
Gas concentration expressed in percent volume units.
Gas Type
Precision / Accuracy
CO
2
±(0.3 vol% + 4% reading)
Note 1:
Specification for precision is valid for any environmental condition specified, except in the cases expressed in the table below with
“Effects of Gas and Steam Interference”.
Table 17: Interference specifications in the Capnography
Effects of Gas and Steam Interference:
Gases or Steam
Gas Level
CO
2
N
2
O
60 vol%
---
(1 and 2)
HAL
4 vol%
---
(1)
ENF, ISO, SEV
5 vol%
+8% measurement read..
(3)
DES
15 vol%
+12% measurement read.
(3)