33
OxyMag_rev26
Test Sequence
The tests are essential to check if the equipment is operating as expected and make adjustments for the best possible
performance. Remember to conduct initial tests before starting ventilation.
•
Test Sequence must be performed with the patient disconnected.
1.
Home screen - Press the Test button and the sequence of internal tests will be activated. Follow the instructions on the
screen.
2.
Upon entering the home screen of the test sequence, you should hear a sequence of “beeps” in conjunction with the
lighting of the light alarm indicator. If you do not hear the audible signal or do not see the light signal above the liquid
Crystal display, press key NO; otherwise press YES to proceed to the next test.
3.
B
y pressing the key ”NO”, the message: “Inoperative Device” –
Contact Technical assistance will appear. The equipment
requests the clearance of y-connector and the sensor connection corresponding to the last patient ventilated. Press Ok
when this condition is carried out.
•
After a ventilation, to change the patient type in the test sequence, restart the equipment,
select the type of patient desired and restart the equipment again, only then proceed to the
test sequence.
4.
Tests will be carried out sequentially; after each item, there is a report of pass (OK message) or fail (Failure message).
•
If any test shows
Failure
perform the required repair (see table 13)
5.
After the test phase of the proximal sensor, press NEXT to continue.
6.
Occlusion of the breathing circuit in the “Y” after the flow sensor will be requested. Press OK to confirm that the
circuit is
properly occluded.
Make sure that all the test items are APPROVED and check if the data for compliance, breathing circuit resistance and
leakage value are suitable for use in the ventilator.
7.
Press key END to complete.
8.
The system will automatically return to the home screen of the ventilator. From this point on, proceed with the normal
startup of the ventilator.