86
OxyMag_rev26
11.5
Replacing the Ambient Air Filter
To replace the ambient air filter, follow the procedure below:
Figure 28: Example for exchanging air filter
(1)
Remove the filter cover on the left side of the
ventilator, item 2 of the figure.
(2)
Remove the old filter, item 1 of the figure.
(3)
Clean the seating area of the filter with a cotton
ball soaked in water and mild soap solution.
Caution
•
Do not use compressed air for cleaning, as this
may introduce dust and dirt in the gas mixing
system.
(4)
After drying, introduce a new filter.
(5)
Install the filter cover and check if the set is firmly
closed.
•
Use only filters, parts, pieces and
accessories specified by MAGNAMED
listed in this manual, which have been
tested and approved for use in
conjunction with this equipment;
otherwise, this can jeopardize the
operation endangering the patient or
user;
•
filter when saturated generates an
increase in the resistance of the ambient
air inflow and can make the minimum
concentrations (35% O2) to not be met.
In this case, replace the filter.
Caution
•
Do not operate the equipment without this filter, because it may damage the system controlling the air/oxygen
mixture.
11.6
Forwarding the Product to Repair Service
Products before being sent to repair service should be cleaned and disinfected as directed in this manual Cleanning and Sterilization
(capítulo 11). Products showing signs of potential hospital contaminants will be returned without repair service in order to be disinfected
prior to the service.