94
MD 610_1c 09/2015
COD HR
with Vario Tube Test
0 – 15 g/l O
2
(=
^
0 – 15 000 mg/l O
2
)
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add
0.2 ml
deionised water
(this is the blank (Note 1)).
2. Open another white capped reaction vial and add
0.2 ml of the water sample
(this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently
several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for
120 minutes
in the preheated reactor
at a temperature of
150°C
.
5.
(CAUTION: The vials are hot!)
Remove the tubes from the heating block and allow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are
l
aligned.
7. Press
ZERO
key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are
l
aligned.
10. Press
TEST
key.
The result is shown in the display in
g/l
COD.
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Ø 16 mm
2
3
1
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...