81
MD 610_1c 09/2015
10. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
11. Place the vial in the sample chamber making sure that
the marks are aligned.
12. Press
TEST
key.
Record the displayed test result (G).
13. Remove the vial from the sample chamber, empty the
vial, rinse vial and cap several times. Fill with
a few
drops of water sample
.
14. Add
one DPD No. 1 tablet
straight from the foil and
crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
17. Place the vial in the sample chamber making sure that
the marks are aligned.
18. Press
TEST
key.
Record the displayed test result (A).
19. Remove the vial from the sample chamber.
20. Add
one DPD No. 3 tablet
straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
21. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that
the marks are aligned.
Zero accepted
prepare Test
press TEST
Zero accepted
prepare Test
press TEST
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...