245
MD 610_1c 09/2015
Notes:
1. For photometric determination of pH values only use THYMOLBLUE tablets in black
printed foil pack and marked with PHOTOMETER.
2. pH values below 8.0 and above 9.6 can produce results inside the measuring range.
A plausibility test (pH-meter) is recommended.
3. The accuracy of the colorimetric determination of pH values depends on various
boundary conditions (buffer capacity of the sample, salt contents etc.).
4. Salt error
Correction of test results (average values) for samples with salt contents of:
Indicator
Salt content
Thymolblue
1 molar
– 0.22
2 molar
– 0.29
3 molar
– 0.34
The values of Parson and Douglas (1926) are based on the use of Clark and Lubs buffers.
1 Mol NaCl = 58.4 g/l = 5.8 %
Reagent
Form of reagent/Quantity
Order-No.
THYMOLBLUE PHOTOMETER
Tablet / 100
515710