Phosphate LR
with Liquid reagent
0.1 – 10 mg/l PO
4
This test is suitable for determining ortho-Phosphate in
boiler waters and potable water supplies. Samples should be
filtered prior to analysis to remove any suspended insoluble
phosphate. A GF/C filter is suitable.
Unscrew the two halves of the filter holder and place one
GF/C filter circle onto the base section. Screw the two
parts together again,
ensuring the O ring is correctly
located
.
1. Fill a clean 20 ml syringe with approx. 14 ml water
sample.
2. Connect the syringe to the filtration assembly and
discharge the syringe to waste, down to the 10 ml
mark.
3. Fill a clean vial (24 mm Ø) with
10 ml of water sample
from the prepared syringe
, close tightly with the
cap.
4. Place the vial in the sample chamber making sure that
the marks are aligned.
5. Press
ZERO
key.
6. Remove the vial from the sample chamber.
7. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
50 drops KS80 (CRP)
8. Close the vial tightly with the cap and invert several
times to mix the contents.
9. Add
one level spoon of reagent KP119 (Ascorbic
Acid)
to the same water sample (note 1).
3
4
3
prepare Zero
press ZERO
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...