Silica LR / Silicon dioxide LR
with Vario Powder Pack
and Liquid Reagent
0.1 – 1.6 mg/l SiO
2
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill each vial with
10 ml of the water sample
.
2. Add
0.5 ml Vario Molybdate 3 reagent solution
into
each vial.
3. Close the vials tightly with the caps and swirl several
times to mix the contents (Note 1).
4. Press
[ ]
key.
Wait for a
reaction period of 4 minutes
(Note 2).
After the reaction period is finished proceed as follows:
5. Add the contents of
one Vario Silica Citric Acid F10
Powder Pack
straight from the foil into each vial.
6. Close the vials tightly with the caps and swirl several
times to mix the contents.
7. Press
[ ]
key.
Wait for a
reaction period of 1 minute
(Note 3).
After the reaction period is finished proceed as follows:
8. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
9. Add the contents of
one Vario LR Silica Amino Acid
F F10 Powder Pack
straight from the foil into the vial
(the sample).
10. Close the vial tightly with the cap and swirl several times
to mix the contents.
11. Press
ZERO
key (blank is already placed in the sample
chamber – see point 8).
Wait for a
reaction period of 2 minutes
.
Countdown
4:00
start:
Countdown
1:00
start:
1
2
3
4
5
1
3
Countdown
2:00
prepare Zero
press ZERO
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...