136
MD 610_1c 09/2015
Hydrazine
with Vario Liquid Reagent
0.005 – 0.6 mg/l N
2
H
4
/ 5 – 600 μg/l N
2
H
4
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill a clean vial with
10 ml deionised water
(this is the
blank).
2. Add
1 ml VARIO Hydra 2 Rgt Solution
into the vial
(Note 3).
3. Close the vial tightly with the cap and swirl several times
to mix the contents.
4. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
5. Press
ZERO
key.
6. Remove the vial from the sample chamber.
7. Fill the second clean vial with
10 ml of the water
sample
(this is the sample).
8. Add
1 ml VARIO Hydra 2 Rgt Solution
into the vial.
9. Close the vial tightly with the cap and swirl several times
to mix the contents.
10. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
11. Press
TEST
key.
Wait for a
reaction period of 12 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as Hydrazine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
12:00
0
6
2
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...