Notes:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%).
Rinse them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used:
Fluoride
[mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05
0.10
0.15
0.20
0.25
0.30
0.2
0.05
0.11
0.16
0.21
0.27
0.32
0.4
0.06
0.11
0.17
0.23
0.28
0.34
0.6
0.06
0.12
0.18
0.24
0.30
0.37
0.8
0.06
0.13
0.20
0.26
0.32
0.40
1.0
0.07
0.13
0.21
0.28
0.36
0.45
1.5
0.09
0.20
0.29
0.37
0.48
---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. A special tablet ingredient prevents effects on the measurement due to iron and
manganese.
5. Al
Al
2
O
3
Reagent
Form of reagent/Quantity
Order-No.
Set
ALUMINIUM No. 1 / No. 2
Tablet / per 100
inclusive stirring rod
517601BT
ALUMINIUM No. 1
Tablet / 100
515460BT
ALUMINIUM No. 2
Tablet / 100
515470BT
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...