60
MD 610_1c 09/2015
Chlorine, free
with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add the contents of
one VARIO Chlorine FREE-DPD /
F10 Powder Pack
straight from the foil to the water
sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the
marks are aligned.
8. Press
TEST
key.
The result is shown in the display in mg/l free
Chlorine.
Notes:
See page 47
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
1
2
3
4
0
1
1
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...