193
MD 610_1c 09/2015
11. Close the vials with the caps and shake to mix the
contents (at least 15 seconds, Note 8).
12. Press
[ ]
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished proceed as follows:
13. Open
two TN Acid LR/HR (Reagent C) vials
and
add
2 ml of the digested, treated blank
to one vial
(this is the blank).
14. Add
2 ml of the digested, treated water sample
to
the other TN Acid LR/HR vial (this is the sample).
15. Close the vials with the caps and swirl the vials gently
several times to mix the contents (10 x, Note 9).
(CAUTION: Vials warm up)
.
16. Place the vial (the blank) in the sample chamber making
sure that the
l
marks are aligned.
17. Press
ZERO
key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished the measurement
starts automatically.
18. Remove the vial from the sample chamber.
19. Place the vial (the sample, Note 10) in the sample
chamber making sure that the
l
marks are aligned.
20. Press
TEST
key.
The result is shown in the display in mg/l Nitrogen.
Notes and Reagent:
see page 194
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
Countdown
2:00
start:
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...