Recall results of a selected Code No. range
Press [MODE], [Shift] + [3][2] keys.
Confirm with [ ] key.
The display shows:
Enter numeric code number (up to 6 places) for the first
required Code No., e.g.: [Shift] + [1].
Confirm with [ ] key.
The display shows:
Enter numeric code number (up to 6 places) for the last
required Code No., e.g.: [Shift] + [1][0].
Confirm with [ ] key.
The display shows:
• Press
[ ] key and all stored results in the selected Code
No. range are displayed.
• Press [F3] key to transfer the displayed result.
• Press [F2] key to transfer all selected results.
• End with [ESC].
Note:
1. It is possible to cancel the entry by [ESC].
2. If you want to recall only results of one code number enter the same code number
twice.
3. If you want to recall all results without code no. (code no. is 0) enter Zero [0] twice.
<Storage>
sorted: Code-No.
from _ _ _ _ _ _
to _ _ _ _ _ _
from
000001
to
000010
Start: cancel: ESC
Transf., single: F3
Transf., all: F2
Mode
2
3