321
MD 610_1c 09/2015
Fluoride Method 170
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [3] keys.
The display shows:
1. Fill a clean vial (24 mm Ø) with exactly
10 ml of deionised
water
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
exactly 2 ml SPADNS reagent solution
to the
water sample.
Caution: Vial is filled up to the top!
6. Close the vial tightly with the cap and swirl gently
several times to mix the contents.
<Calibration>
M170 Fluoride L
Zero: deionised water
press ZERO
<Calibration>
1: M191 Ca-Hardness 2
2: M191 0 Jus. Reset
3: M170 Fluoride L
Press [Shift] + [0] keys to keep the method blank.
Press [Shift] + [1] keys to erase the method blank and set
the value back to factory calibration.
The instrument goes back to mode menu automatically.
Mode
0
4
0
1
Shift
Shift
3
Shift