182
MD 610_1c 09/2015
Nitrate
with Tablet and Powder
0.08 – 1 mg/l N
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber and empty
the vial.
5. Fill the Nitrate test tube with
20 ml of the water
sample
.
6. Add
1 level spoon of Nitrate Test powder
.
7. Close the tube tightly with the cap and swirl vigorously
for one minute.
8. Add
one NITRATE TEST tablet
straight from the foil
to the water sample.
9. Close the tube tightly with the cap and swirl vigorously
for one minute.
10. Stand the tube upright and after the reducing agent
has settled to the bottom, gently invert it three to four
times so as to complete the flocculation of the reducing
agent. Then let the tube stand for a further 2 minutes.
Open the tube and wipe around the top of the tube with
a clean tissue to remove any residuals of the reducing
agent.
11. Carefully decant 10 ml of the treated solution into
the vial (24 mm Ø) used for zeroing, ensuring that no
reducing agent is carried over.
12. Add
one NITRITE LR tablet
straight from the foil to the
water sample and crush the tablet using a clean stirring
rod.
6
0
2
prepare Zero
press ZERO
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...