246
MD 610_1c 09/2015
Polyacrylate
with Liquid reagent
1 – 30 mg/l
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
1 ml (25 drops) KS255 (Polyacrylate reagent 1)
to the water sample (note 1).
6. Close the vial tightly with the cap and swirl gently several
times.
7. Add
1 ml (25 drops) KS256 (Polyacrylate reagent 2)
to the water sample (note 1).
8. Close the vial tightly with the cap and swirl gently several
times.
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
Wait for a
reaction period of 10 minutes
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Polyacrylic Acid
2'100 sodium salt.
3
8
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00