Interference levels decrease with increasing sample volume.
Example: Iron interferes above 200 mg/L if a sample volume of 5 ml is used.
At a sample volume of 10 ml the interference level decreases to 100 mg/L.
Table 3:
Interfering substances
Interference level using 5 ml of sample
Aluminium
100 mg/l
Arsenate
interferes at all concentrations
Benzotriazole
10 mg/l
Bicarbonate
1000 mg/l
Bromide
100 mg/l
Calcium
5000 mg/l
CDTA
100 mg/l
Chloride
5000 mg/l
Chromate
100 mg/l
Copper
100 mg/l
Cyanide
100 mg/l; increase the UV digestion
to 30 minutes
Diethanoldithiocarbamate
50 mg/l
EDTA
100 mg/l
Iron
200 mg/l
Nitrate
200 mg/l
NTA
250 mg/l
ortho-Phosphate
15 mg/l
Phosphite and
organophosphorus compounds
reacts quantitatively;
Meta- and Polyphosphates do not interfere
Silica
500 mg/l
Silicate
100 mg/l
Sulfate
2000 mg/l
Sulfide
interferes at all concentrations
Sulfite
100 mg/l
Thiourea
10 mg/l
highly buffered samples or
extreme sample pH
may exceed the buffering capacity of the
reagents and require sample pretreatment
Summary of Contents for MD 610
Page 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Page 2: ......
Page 4: ......
Page 16: ...8 MD 610_1c 09 2015 ...
Page 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Page 23: ...15 MD 610_1c 09 2015 ...
Page 85: ...77 MD 610_1c 09 2015 ...
Page 93: ...85 MD 610_1c 09 2015 ...
Page 294: ...286 MD 610_1c 09 2015 ...
Page 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Page 298: ...290 MD 610_1c 09 2015 ...
Page 350: ...342 MD 610_1c 09 2015 ...