Gocator Line Profile Sensors - User Manual
Tools and Native Drivers • GenTL Driver • 441
7.
Click OK in the dialogs until they are all closed.
To work with the Gocator GenTL driver, the Gocator must operate in Surface or Video mode with its the
appropriate output enabled in the
Ethernet
panel in the
Output
page. Check
Acquire Intensity
in the
Scan Mode
panel on the
Scan
page and enable intensity output in the
Ethernet
panel if intensity data
is required.
The GenTL driver packs the part output, intensity, and stamps (e.g., time stamp, encoder index, etc.) into
either a 16-bit RGB image or a 16-bit grey scale image. You can select the format in the Go2GenTL.xml
setting file.
The width and height of the 16-bit RGB or grey scale image is calculated from the maximum number of
columns and rows needed to accommodate the sensor's field of view and the maximum part length.
16-bit RGB Image
When the 16-bit RGB format is used, the height map, intensity, and stamps are stored in the red, green,
and blue channel respectively.
Channel
Details
Red
Height map information. The width and height of the image represent the dimensions
in the X and Y axis. Together with the pixel value, each red pixel presents a 3D point in
the real-world coordinates.
The following formula can be used to calculate the real-world coordinates (X, Y, Z) from
pixel coordinates (Px, Py, Pz):
X = X Px * X resolution
Y = Y Py * Y resolution
Z = Z Pz * Z resolution
Refer to the blue channel on how to retrieve the offset and resolution values. If Pz is 0 if
the data is invalid. The Z offset is fixed to -32768 * Z resolution. Z is zero if Pz is 32768.
Green
Intensity information. Same as the red channel, the width and height of the image
represent the dimension in the X and the Y axis. Together with the pixel value, each
blue pixel represents an intensity value in the real-world coordinates.
The following formula can be used to calculate the real-world coordinates (X, Y, Z) from
pixel coordinates (Px, Py, Pz):
X = X Px * X resolution
Y = Y Py * Y resolution
Z = 16-bit intensity value
The intensity value is 0 if the intensity image is not available. Gocator outputs 8-bit
intensity values. The values stored in the 16-bit RGB image is multiplied by 256. To
obtain the original values, divide the intensity values by 256.
Refer to the blue channel on how to retrieve the offset and resolution values.
Blue
Stamp information. Stamps are 64-bit auxiliary information related to the height map
and intensity content. The next table explains how the stamps are packed into the blue
Summary of Contents for Gocator 2100 Series
Page 197: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 197 ...
Page 215: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 215 ...
Page 462: ...Gocator Line Profile Sensors User Manual Specifications Sensors 462 Dimensions Envelope ...
Page 464: ...Gocator Line Profile Sensors User Manual Specifications Sensors 464 Envelope ...
Page 466: ...Gocator Line Profile Sensors User Manual Specifications Sensors 466 Dimensions Envelope ...
Page 468: ...Gocator Line Profile Sensors User Manual Specifications Sensors 468 Dimensions Envelope ...
Page 470: ...Gocator Line Profile Sensors User Manual Specifications Sensors 470 Dimensions ...
Page 471: ...Gocator Line Profile Sensors User Manual Specifications Sensors 471 Envelope ...
Page 473: ...Gocator Line Profile Sensors User Manual Specifications Sensors 473 Dimensions ...
Page 474: ...Gocator Line Profile Sensors User Manual Specifications Sensors 474 Envelope ...
Page 476: ...Gocator Line Profile Sensors User Manual Specifications Sensors 476 Dimensions ...
Page 477: ...Gocator Line Profile Sensors User Manual Specifications Sensors 477 Envelope ...
Page 479: ...Gocator Line Profile Sensors User Manual Specifications Sensors 479 Dimensions ...
Page 480: ...Gocator Line Profile Sensors User Manual Specifications Sensors 480 Envelope ...
Page 484: ...Gocator Line Profile Sensors User Manual Specifications Sensors 484 Dimensions ...
Page 485: ...Gocator Line Profile Sensors User Manual Specifications Sensors 485 Envelope ...
Page 487: ...Gocator Line Profile Sensors User Manual Specifications Sensors 487 Dimensions ...
Page 488: ...Gocator Line Profile Sensors User Manual Specifications Sensors 488 Envelope ...
Page 491: ...Gocator Line Profile Sensors User Manual Specifications Sensors 491 Dimensions ...
Page 492: ...Gocator Line Profile Sensors User Manual Specifications Sensors 492 Envelope ...
Page 506: ...Gocator Line Profile Sensors User Manual Specifications Master Hubs 506 Master 810 Dimensions ...