Parameter
Description
Location
(Strip Height, Strip X, and
Strip Z measurements only)
Specifies the strip position from which the measurements are performed.
Left
- Left edge of the strip.
Right
- Right edge of the strip.
Center
- Center of the strip.
Left Edge
Right Edge
Specifies the features that will be considered as the strip's left and right edges. You can select
more than one condition.
Rising
- Rising edge detected based on the strip edge parameters.
Falling
- Falling edge detected based on the strip edge parameters.
Data end
- First valid profile data point in the measurement region.
Void
- Gap in the data that is larger than the maximum void threshold. Gaps connected to the
measurement region's boundary are not considered as a void.
See "Strip Start and Terminate Conditions" in the
Gocator Measurement Tool Technical Manual
for
the definitions of these conditions.
Select Type
Specifies how a strip is selected when there are multiple strips within the measurement area.
Best
- The widest strip.
Index Left
- 0-based strip index, counting from left to right.
Index Right
- 0-based strip index, counting from right to left.
Index
0-based strip index.
Min Height
Specifies the minimum deviation from the strip base. See "Strip Step Edge Definitions" in the
Gocator Measurement Tool Technical Manual
on how this parameter is used for different base
types.
Support Width
Specifies the width of the region around the edges from which the data is used to calculate the
step change. See "Strip Step Edge Definitions" in the
Gocator Measurement Tool Technical Manual
on how this parameter is used by different base types.
Transition Width
Specifies the nominal width needed to make the transition from the base to the strip. See
"Strip Step Edge Definitions" in the
Gocator Measurement Tool Technical Manual
on how this
parameter is used by different base types.
Max Void Width
The maximum width of missing data allowed for the data to be considered as part of a strip
when
Void
is selected in the
Left
or
Right
parameter. This value must be smaller than the edge
Support Width
.
When occlusion and exposure causes data drops, users should use the gap filling function to
fill the gaps. See
on page 105 for information.
Summary of Contents for Gocator 2100 Series
Page 197: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 197 ...
Page 215: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 215 ...
Page 462: ...Gocator Line Profile Sensors User Manual Specifications Sensors 462 Dimensions Envelope ...
Page 464: ...Gocator Line Profile Sensors User Manual Specifications Sensors 464 Envelope ...
Page 466: ...Gocator Line Profile Sensors User Manual Specifications Sensors 466 Dimensions Envelope ...
Page 468: ...Gocator Line Profile Sensors User Manual Specifications Sensors 468 Dimensions Envelope ...
Page 470: ...Gocator Line Profile Sensors User Manual Specifications Sensors 470 Dimensions ...
Page 471: ...Gocator Line Profile Sensors User Manual Specifications Sensors 471 Envelope ...
Page 473: ...Gocator Line Profile Sensors User Manual Specifications Sensors 473 Dimensions ...
Page 474: ...Gocator Line Profile Sensors User Manual Specifications Sensors 474 Envelope ...
Page 476: ...Gocator Line Profile Sensors User Manual Specifications Sensors 476 Dimensions ...
Page 477: ...Gocator Line Profile Sensors User Manual Specifications Sensors 477 Envelope ...
Page 479: ...Gocator Line Profile Sensors User Manual Specifications Sensors 479 Dimensions ...
Page 480: ...Gocator Line Profile Sensors User Manual Specifications Sensors 480 Envelope ...
Page 484: ...Gocator Line Profile Sensors User Manual Specifications Sensors 484 Dimensions ...
Page 485: ...Gocator Line Profile Sensors User Manual Specifications Sensors 485 Envelope ...
Page 487: ...Gocator Line Profile Sensors User Manual Specifications Sensors 487 Dimensions ...
Page 488: ...Gocator Line Profile Sensors User Manual Specifications Sensors 488 Envelope ...
Page 491: ...Gocator Line Profile Sensors User Manual Specifications Sensors 491 Dimensions ...
Page 492: ...Gocator Line Profile Sensors User Manual Specifications Sensors 492 Envelope ...
Page 506: ...Gocator Line Profile Sensors User Manual Specifications Master Hubs 506 Master 810 Dimensions ...