Gocator Line Profile Sensors - User Manual
Gocator Web Interface • Scan Setup and Alignment • 105
Clearing Alignment
Alignment can be cleared to revert the sensor to sensor coordinates.
To clear alignment:
1.
Go to the
Scan
page.
2.
Choose Profile or Surface mode in the
Scan Mode
panel, depending on the type of measurement
whose decision you need to configure.
If one of these modes is not selected, tools will not be available in the
Measure
panel.
3.
Expand the
Alignment
panel by clicking on the panel header or the
button.
4.
Click the
Clear Alignment
button.
The alignment will be erased and sensors will revert to using sensor coordinates.
Filters
Filters are used to post-process scan data along the X or Y axis to remove noise or clean it up before it is
output or is used by measurement tools.
The following types of filters are supported:
Filter
Description
Gap Filling
Fills in missing data caused by occlusions using information from the nearest neighbors.
Gap filling also fills gaps where no data is detected, which can be due to the surface
reflectivity, for example dark or specular surface areas, or to actual gaps in the surface.
Median
Substitutes the value of a data point with the median within a specified window around
the data point.
Smoothing
Applies moving window averaging to reduce random noise.
Decimation
Reduces the number of data points.
Filters are applied in the order displayed in the table above. The filters are configured in the
Filters
panel
on the
Scan
page.
Gap Filling
Gap filling works by filling in missing data points using either the lowest values from the nearest
neighbors or linear interpolation between neighboring values (depending on the Z difference between
neighboring values), in a specified X or Y window. The sensor can fill gaps along both the X axis and the Y
Summary of Contents for Gocator 2100 Series
Page 197: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 197 ...
Page 215: ...Gocator Line Profile Sensors User Manual Gocator Web Interface Measurement 215 ...
Page 462: ...Gocator Line Profile Sensors User Manual Specifications Sensors 462 Dimensions Envelope ...
Page 464: ...Gocator Line Profile Sensors User Manual Specifications Sensors 464 Envelope ...
Page 466: ...Gocator Line Profile Sensors User Manual Specifications Sensors 466 Dimensions Envelope ...
Page 468: ...Gocator Line Profile Sensors User Manual Specifications Sensors 468 Dimensions Envelope ...
Page 470: ...Gocator Line Profile Sensors User Manual Specifications Sensors 470 Dimensions ...
Page 471: ...Gocator Line Profile Sensors User Manual Specifications Sensors 471 Envelope ...
Page 473: ...Gocator Line Profile Sensors User Manual Specifications Sensors 473 Dimensions ...
Page 474: ...Gocator Line Profile Sensors User Manual Specifications Sensors 474 Envelope ...
Page 476: ...Gocator Line Profile Sensors User Manual Specifications Sensors 476 Dimensions ...
Page 477: ...Gocator Line Profile Sensors User Manual Specifications Sensors 477 Envelope ...
Page 479: ...Gocator Line Profile Sensors User Manual Specifications Sensors 479 Dimensions ...
Page 480: ...Gocator Line Profile Sensors User Manual Specifications Sensors 480 Envelope ...
Page 484: ...Gocator Line Profile Sensors User Manual Specifications Sensors 484 Dimensions ...
Page 485: ...Gocator Line Profile Sensors User Manual Specifications Sensors 485 Envelope ...
Page 487: ...Gocator Line Profile Sensors User Manual Specifications Sensors 487 Dimensions ...
Page 488: ...Gocator Line Profile Sensors User Manual Specifications Sensors 488 Envelope ...
Page 491: ...Gocator Line Profile Sensors User Manual Specifications Sensors 491 Dimensions ...
Page 492: ...Gocator Line Profile Sensors User Manual Specifications Sensors 492 Envelope ...
Page 506: ...Gocator Line Profile Sensors User Manual Specifications Master Hubs 506 Master 810 Dimensions ...