9. EYE PATTERN DISPLAY (SER02/SER09)
195
9.7.3
Selecting the Filter
To select the filter that is used during jitter measurement, follow the procedure below. The
selected filter is indicated in the lower right of the display.
If you change this setting, the filter that you selected for jitter display also changes.
Reference Section 9.8.3, “Selecting the Filter”
Procedure
EYE
→
F•3
EYE SETUP →
F•3 FILTER: 100kHz / 1kHz / 100Hz / 10Hz / TIMING /
ALIGNMENT
Settings
100kHz:
Jitter at 100 kHz or higher is measured.
1kHz:
Jitter at 1 kHz or higher is measured.
100Hz:
Jitter at 100 Hz or higher is measured.
10Hz:
Jitter at 10 Hz or higher is measured.
TIMING:
Timing jitter is measured. Jitter at 10 Hz or higher is measured.
ALIGNMENT: Alignment jitter is measured. When the input signal is not SD, jitter at 100
kHz and higher is measured. When the input signal is SD, jitter at 1 kHz
and higher is measured.