72
c
HAPTER
4:
Computer Interface Operation
MeasureReady™ M91 FastHall™ Measurement Controller
*SRE
Command Format
*SRE <data>
Parameter
<data>
Decimal value that is the sum of the binary-weighted
values for the desired bits.
Data type is
NR1
Query Format
*SRE?
Return Parameter
<data>
(see above)
Examples
Command
*SRE 4
Enables error available flag
Query
*SRE?
Query Response
136
Indicates that operation and questionable event enable
flags are set
Remarks
This command programs the service request enable register bits. This register deter-
mines which bits from the status byte register are summed to set the Master Status
Summary (MSS) bit. To enable a status flag bit, send the command *SRE with the sum
of the bit weighting for each desired bit. Refer to section 4.3.2.2 for a list of status
flags.
Bit
Bit weighting
Event name
2
4
EAV
3
8
QSB
4
16
MAV
5
32
ESB
7
128
OSB
Total:
188
*STB?
Query Format
*STB?
Return Parameter
<data>
Data type is
NR1
Examples
Query
*OPC?
Query Response
1
Remarks
Reads the status byte register. Acts like a serial poll, but does not reset the register to
all zeros. The integer returned represents the sum of the bit weighting of the status
flag bits that are set in the Status Byte Register. Refer to section 4.3.2.1 for a list of sta-
tus flags.
*TST?
Query Format
*TST?
Return Parameter
<status>
0 = no errors found, 1= errors found
Data type is
NR1
Examples
Query
*TST?
Query Response
0
Indicates no errors were found
Remarks
The report status based on test done at power up..
*WAI
Command Format
*WAI
Remarks
Causes the interface to hold off until all pending operations have been completed.
This is the same function as the *OPC command, except that it does not set the opera-
tion complete event bit in the event status register.
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...