![Lake Shore Cryotronics Measure Ready M91 FastHall User Manual Download Page 72](http://html.mh-extra.com/html/lake-shore-cryotronics/measure-ready-m91-fasthall/measure-ready-m91-fasthall_user-manual_670649072.webp)
FIGURE 4-3
Standard event status register
OR
AND
AND
AND
AND
AND
AND
Not
used
Not
used
PON
CME
EXE
DSE
QYE
OPC
7
6
5
4
3
2
1
0
Not
used
Not
used
PON
CME
EXE
DSE
QYE
OPC
7
6
5
4
3
2
1
0
Standard event
status register
*ESR?
Standard event status
enable register
*ESE, *ESE?
— Bit
— Name
— Bit
— Name
To event
summary
bit (ESB)
of status
byte register
62
c
HAPTER
4:
Computer Interface Operation
MeasureReady™ M91 FastHall™ Measurement Controller
4.3.2.4 Operation Event Register Set
The operation event register reports the instrument events that are considered part
of normal operation. Any or all of these events may be reported in the operation event
summary bit through the enable register (FIGURE 4-4.) The operation event enable
command (
STATus:OPERation:ENABle
) programs the enable register and the
query command (
STATus:OPERation:ENABle?
) reads it.
STATus:OPERation[:EVENt]?
reads and clears the operation event register.
STATus:OPERation:CONDition?
reads the operation condition register.
The used bits of the operation event register are described as follows:
D
Measuring Done (MEAS), Bit (4):
this bit is set when a measurement has finished run-
ning. Reset once a new measurement has started.
FIGURE 4-4
Operation event status register
To operation
event summary
bit (OSB) of
status byte
register
AND
OR
AND
AND
AND
Not
used
Not
used
Not
used
Not
used
15 14 13 12 11 10 9
8
Not
used
Not
used
Not
used
Not
used
MEAS
Not
used
Not
used
Not
used
7
6
5
4
3
2
1
0
Not
used
Not
used
Not
used
Not
used
15 14 13 12 11 10 9
8
Not
used
Not
used
Not
used
Not
used
MEAS
Not
used
Not
used
Not
used
7
6
5
4
3
2
1
0
Not
used
Not
used
Not
used
Not
used
15 14 13 12 11 10 9
8
Not
used
Not
used
Not
used
Not
used
MEAS
Not
used
Not
used
Not
used
7
6
5
4
3
2
1
0
Operation Event Enable Register
STATus:OPERation:ENABle
STATus:OPERation:ENABle?
Operation Event Register
STATus:OPERation[:EVENt]?
Operation Condition Register
STATus:OPERation:CONDition?
— Bit
— Name
— Bit
— Name
— Bit
— Name
=
MEAS
Measuring done
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Not
used
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...