112
c
HAPTER
4:
Computer Interface Operation
MeasureReady™ M91 FastHall™ Measurement Controller
HALl:HBAR[:DC]:STARt
Summary
Performs a DC Hall measurement for a Hall bar sample.
Command Format
HALl:HBAR[:DC]:STARt
<excitationType>,<excitationValue>,<excitationRange>,
<excitationMeasurementRange>,<measurementRange>,<complianceLimit>,
<maxNumberOfSamples>,<userDefinedFieldReadingInTesla>,
<withFieldReversal>,<resistivity>,<blankingTime>,<sampleThickness>,
<minimumSnr>,<samplingTime>
Parameters
<excitationType> VOLTage,
CURRent
Data type is
NAMED
<excitationValue>
For voltage excitation, -10 to 10 V
For current excitation, -100 e-3 to 100 e-3 A
Data type is
NRf
<excitationRange>
For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<excitationMeasurementRange> For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the measurement range to the best fit range for a
given excitation range.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<measurementRange>
For voltage excitation, specify the current measurement
range 0 to 100 e-3 A
For current excitation, specify the voltage measurement
range 0 to 10 V
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<complianceLimit>
For voltage excitation, specify the current limit 100 e-9 to
100 e-3 A
For current excitation, specify the voltage compliance
1 to 10 V
Data type is
NRf
<maxNumberOfSamples> When minimumSnr is omitted or INFinity (9.9 e+37),
the total number of samples to average is 1-1000.
When minimumSnr is specified, the maximum number of
samples to average is 10-1000.
Data type is
Number
<userDefinedFieldReadingsInTesla>The field, in units of tesla, the sample is being
subjected to. Used for calculations.
Data type is
NRf
<withFieldReversal>
Specifies whether or not to apply reversal field.
DEFault = true
Data type is
bool
<resistivity>
The resistivity of the sample in units of ohm*meters (bulk)
of ohms per square (sheet). Used for calculations.
Measure this value using the RESistivity SCPI subsystem.
Defaults to not a number which will propagate through
calculated values.
DEFault = NaN (9.91 e+37)
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...