126
c
HAPTER
4:
Computer Interface Operation
MeasureReady™ M91 FastHall™ Measurement Controller
100 e-3 A
For current excitation, specify the voltage compliance
1 to 10 V
Data type is
NRf
<maxNumberOfSamples> When minimumSnr is omitted or INFinity (9.9 e+37),
the total number of samples to average is 1-1000.
When minimumSnr is specified, the maximum number of
samples to average is 10-1000.
Data type is
Number
<userDefinedFieldReadingInTesla> The field, in units of tesla, the sample is being
subjected to. Used for calculations.
Data type is
NRf
<withFieldReversal>
Specifies whether or not to apply reversal field.
DEFault = true
Data type is
bool
<resistivity>
The resistivity of the sample in units of ohm*meters (bulk)
of ohms per square (sheet). Used for calculations. Measure
this value using the RESistivity SCPI subsystem. Defaults to
not a number which will propagate through calculated
values.
DEFault = NaN (9.91 e+37)
NOTE:
When the resistivity is omitted or NaN (9.91 e+37),
mobility is not calculated.
Data type is
Number
<blankingTime>
The time, in seconds, to wait for hardware to settle before
gathering readings.
0.5 ms to 300 s with a resolution of 0.1 ms
DEFault = 2 ms
MINimum = 0.5 ms
MAXimum = 300 s
Data type is
Number
<sampleThickness>
Thickness of the sample in meters.
0 to 10 e-3 m
DEFault = 0 m
Data type is
Number
<MinimumSnr>
The desired signal-to-noise ratio of the measured
resistance, calculated using measurement average voltage/
error of mean 1 - 1000, or INFinity (9.9e+37).
DEFault = Infinity
Data type is
Number
<samplingTime>
The sampling time, in seconds, which measurements will be
averaged over to get one sample.
MINimum = 0.010 ms
MAXimum = 1 s
DEFault = 1/60 s
Data type is
Number
Example
HAL:DC:START Current,10E-3,10E-3,10E-3,1.0,10,5,0.6,1,0.216,100e-3,1e-2,30
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...