"IsRunning": false,
"StartTime": "1970-01-02T03:03:42.290236-05:00",
"EndTime": "1970-01-02T03:03:46.064352-05:00",
"DurationInSeconds": 3.773
}
RESistivity:RESult:JSON:DATA?
Summary
Retrieves the resistivity and F value of both geometries A and B along with all mea-
surements for a given sample. Van der Pauw results are shown in this example.
Query Format
RESistivityl:RESult:JSON:DATA? <sampleIndex>,<pretty>
Parameters
<sampleIndex>
Sample index, zero based
Data type is
NR1
<pretty>
Optional, pretty format JSON response. Defaults to False.
Data type is
bool
Returns
JSON serialized resistivity measurement sample data.
Data type is
NAMED
Examples
Query RESistivity:RESult:JSON:DATA? 4,1
Query response
{
"ResistivityInOhmMeters": 0.00083849203674707482,
"SheetResistivityInOhmsPerSquare": 0.83849203674707473,
"GeometryAResistivityInOhmMeters": 0.00089504567597727147,
"GeometryASheetResistivityInOhmsPerSquare": 0.89504567597727147,
"GeometryAFValue": 0.48411577862915228,
"GeometryBResistivityInOhmMeters": 0.00078193839751687817,
"GeometryBSheetResistivityInOhmsPerSquare": 0.7819383975168781,
"GeometryBFValue": 0.43807298616439694,
"Measurements": [
{
"ContactConfiguration": "R2134",
"PositiveExcitation": {
"VoltageInVolts": 7.875055E-05,
"CurrentInAmps": 9.994036E-06,
"InCompliance": false,
"VoltageOverload": false,
"CurrentOverload": false
},
"NegativeExcitation": {
"VoltageInVolts": 7.839359E-05,
"CurrentInAmps": -9.994946E-06,
"InCompliance": false,
"VoltageOverload": false,
"CurrentOverload": false
},
"ResistanceInOhms": 0.01785782,
"InCompliance": false,
"VoltageOverload": false,
"CurrentOverload": false
},
{
"ContactConfiguration": "R3241",
"PositiveExcitation": {
"VoltageInVolts": 8.848818E-05,
"CurrentInAmps": 1.000711E-05,
"InCompliance": false,
"VoltageOverload": false,
"CurrentOverload": false
},
"NegativeExcitation": {
"VoltageInVolts": 7.251695E-05,
"CurrentInAmps": -1.000757E-05,
"InCompliance": false,
"VoltageOverload": false,
"CurrentOverload": false
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...