A.5.5 Making a
Measurement
The basic Hall measurement consists of three steps:
1. Checking the sample contacts to ensure they are ohmic (i.e. linear).
2. Measuring the resistivity. These first two steps are performed at 0 T field.
3. Placing the sample into a known magnetic field and measuring the Hall voltage.
After the completion of these measurements, the derived parameters can be calcu-
lated. The FastHall™ station automates the process of entering the required input
parameters, collecting the measurement data, and reporting the output results.
The steps for completing a standard measurement are shown below. For more details,
see
Hall Measurements with MeasureLINK™-MCS Software
, available from the
Downloads page at
.
A.5.5.1 Making Static Hall Measurements with MeasureLINK-MCS™ Software
Scripts for fixed point Hall measurements are included in the MeasureLINK-MCS™
software. This script completes a Hall measurement at a single field for a van der
Pauw sample. To use this script, open the software and follow the steps below.
1. Click
Sequence
.
2. In the Sequence window, click
Measurement
.
3. Click
M91 FastHall™
.
4. Click
Hall measurement
.
5. Fill out the fields in each section of the screen:
D
Preferences:
defines where and how data will be stored, and the units for dis-
playing data.
D
Measurement Setup:
defines the sample parameters of the measurement.
D
Optimization:
determines excitation type and values, and blanking time.
D
Contact Check:
when selected, the measurement will complete a contact
check.
D
Resistivity:
when selected, the measurement will complete a resistivity mea-
surement.
D
Hall:
when selected, the measurement will include a Hall measurement.
Included in the sample card kit is one card with InAs mounted on the card. This is a n-
type material. To run this sample, insert the card into the sample card and mount the
sample card onto the light tight box. Set the MeasureLINK software to these parame-
ters:
D
Preferences:
choose your desired settings, or use the defaults settings.
D
Measurement Setup:
set the sample thickness to 250 microns. Use the default
values for the rest of the settings in this section.
D
Optimization:
click Full optimization, then use the default values.
D
Resistivity:
click Include resistivity, then use the default values.
D
Hall:
click Include Hall, Enter the Magnetic field value on the magnet label. Use
the default values for the rest of the selections.
After setting the parameters, click
Start
. The software will tell you when to move the
magnet. When complete, the carrier type should be n type and the mobility about
24000 cm^2/(V s).
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...