CCHeck:BLANKing:RESult:JSON?
Summary
Returns the results of calculating the optimal blanking time.
Query Format
CCHeck:BLANKing:RESult:JSON? <prettty>
Parameters
<pretty>
Optional, pretty format JSON response. Defaults to False.
Data type is
bool
Returns
JSON serialized contact check optimization blanking time calculation data.
Data type is
NAMED
CCHeck:BLANKing:RUNNing?
Summary
Indicates if the standalone blanking time calculation is running.
Query Format
CCHeck:BLANKing:RUNNing?
Returns
The measurement is running.
Data type is
bool
CCHeck:BLANKing:STARt
Summary
Calculates the optimal blanking time for a contact check on this sample.
Command Format
CCHeck:BLANKing:STARt
CCHeck:HBAR:STARt
Summary
Performs a contact check measurement on contact pairs 5-6, 5-1, 5-2, 5-3, 5-4, and
6-1.
Command Format
CCHeck:HBAR:STARt
<excitationType>,<excitationValueStart>,
<excitationValueEnd>,<excitationRange>,
<measurementRange>,<complianceLimit>,
<numberOfPoints>,<minimumRSquared>,<blankingTime>
<samplingTime>
Parameters
<excitationType> VOLTage,
CURRent
Data type is
ExcitationType
<excitationValueStart>
The starting excitation value.
For voltage excitation -10 to 10 V.
For current excitation -100 e3 to 100 e-3 A.
Data type is
NRf
<excitationValueEnd>
The ending excitation value
For voltage excitation -10 to 10 V
For current excitation -100 e3 to 100 e-3 A
Data type is
NRf
<excitationRange>
For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<measurementRange>
For voltage excitation, specify the current measurement
range 0 to 100 e-3 A.
For current excitation, specify the voltage measurement
range 0 to 10 V.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<complianceLimit>
For voltage excitation, specify the current limit 100 e-9 to
100 e-3 A.
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...