109
|
www.lakeshore.com
FWIRe:RESult[:STANdard][:SUMMary]?
Summary
Retrieves the results of the last run four wire measurement.
Query Format
FWIRe:RESult[:STANdard][:SUMMary]?
Returns
<startDateTime>,<durationInSeconds>,<contactPairPoint1>,<contactPairPoint2>,
<contactPairPoint3>,<contactPairPoint4>,<excitationType>,<excitationValue>,
<excitationRange>,<measurementRange>,<excitationMeasurementRange>,
<complianceLimit>,<blankingTimeInSeconds>,<MaximumNumberOfSamples>,
<MinimumResistanceSnr>,<samplingTimeInSeconds>,<useExcitationReversal>,
<ResistanceAverageInOhms>,<ResistanceStandardErrorInOhms>,
<VoltageAverageInVolts>,<VoltageStandardErrorInAmps>,<CurrentAverageInAmps>,
<CurrentStandardErrorInAmps>,<inCompliance>,<voltageOverload>,
<currentOverload>
<startDateTime>
UTC date and time when the measurement was started in
ISO 8601 format.
Data type is
string
<DurationInSeconds>
Duration of the test in seconds.
Data type is
NR1
<contactPairPoint1>
The contact point assigned to exci.
Data type is
NAMED
<contactPairPoint2>
The contact point assigned to excitation -.
Data type is
NAMED
<contactPairPoint3>
The contact point assigned to voltage measurement
/sense +.
Data type is
NAMED
<contactPairPoint4>
The contact point assigned to voltage measurement/
sense -.
Data type is
NAMED
<excitationType>
The excitation type used for the measurement, VOLTage or
CURRent.
Data type is
NAMED
<excitationValue>
The excitation output value
Data type is
NRf
<excitationRange>
The excitation range.
Data type is
NRf
<measurementRange>
The full scale range of the measurement. For example, if
excitation type is current, this is the full scale voltage range.
Data type is
NRf
<excitationMeasurementRange>
The full scale range used to measure the excitation signal,
dependent on excitationType.
Data type is
NRf
<complianceLimit>
For voltage excitation, the current limit. For current
excitation, the voltage compliance.
Data type is
NRf
<blankingTimeInSeconds>The time in seconds to wait for the hardware to settle
before making measurements.
Data type is
NRf
<MaximumNumberOfSamples> The number of samples to average.
Data type is
NR1
<MinimumResistanceSnr>The desired signal-to-noise ratio of the measured
resistance, calculated using measurement average/
error of mean.
Data type is
NRf
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...