DEFault = 0 m
Data type is
Number
<minimumSnr>
The desired signal to noise ratio of the measurement
calculated using average resistivity / error of mean
1 to1000, or INFinity (9.9 e+37).
DEFault = 30
Data type is
Number
<samplingTime>
The sampling time, in seconds, which measurements will be
averaged over to get one sample.
MINimum = 0.010 ms
MAXimum = 1 s
DEFault = 1/60 s
Data type is
Number
Example
RESistivity:STARt CURRent,10e-6,10e-6,AUTO,100e-3,1.5,25,2.4e-3,0,30
RESisitivity[:VDP]:STARt:LINk
Summary
Performs a resistivity measurement (for a van der Pauw sample) that uses the last run
contact check measurement's excitation type, compliance limit, blanking time, exci-
tation range, and the largest absolute value of the start and end excitation values.
Command Format
RESistivity[:VDP]:STARt:LINk<measurementRange>,<sampleThickness>,
<minimumSnr>,<maxNumberOfSamples>,<samplingTime>
Parameters
<measurementRange>
For voltage excitation, specify the current measurement
range 0 to 100 e-3 A
For current excitation, specify the voltage measurement
range 0 to 10 V
AUTO sets the range to the best fit range
DEFault = AUTO
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<sampleThickness>
Thickness of the sample in meters.
0 to 10 e-3 m
DEFault = 0 m
Data type is
Number
<minimumSnr>
The desired signal-to-noise ratio of the measurement
calculated using average resistivity/error of mean
1 to 1000, or INFinity (9.9 e+37)
DEFault = 30
Data type is
Number
<MaxNumberOfSamples> When minimumSnr is omitted or INFinity (9.9 e+37),
the total numberof samples to average 1 to 1000.
When minimumSnr is specified, the maximum number of
samples to average 10 to 1000.
DEFault = 100
Data type is
Number
<samplingTime>
The sampling time, in seconds, which measurements will be
averaged over to get one sample.
MINimum = 0.010 ms
MAXimum = 1 s
DEFault = 1/60 s
Data type is
Number
Example
RESistivity:STARt:LINk
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...