101
|
www.lakeshore.com
<excitationMeasurementRange>
The full scale range used to measure the excitation signal,
dependent on excitationType.
Data type is
NRf
<measurementRange>
The full scale range of the measurement. For example,
if excitation type is current, this is the full scale voltage
range.
Data type is
NRf
<complianceLimit>
For voltage excitation, the current limit. For current
excitation, the voltage compliance.
Data type is
NRf
<numberOfSamplesToAverage> The number of samples that were taken and averaged
Data type is
NR1
<userDefinedFieldReading> The field, in units of tesla, that the sample is being
subjected to.
Data type is
NRf
<resistivity>
The resistivity of the sample in units of ohm*meters (bulk)
or ohms per square (sheet)
Data type is
NRf
<BlankingTimeInSeconds>The time in seconds to wait for the hardware to settle
before making measurements.
Data type is
NRf
<numberOfVoltageCompensationSamplesToAverage>
The number of voltage compensation samples to average.
Only applied for excitation type voltage.
Data type is
NR1
<SampleThicknessInMeters> Thickness of the sample in meters.
Data type is
NRf
<SamplingTimeInSeconds> The sampling time in seconds.
Data type is
NRf
<minimumSnr>
The desired signal-to-noise ratio of the Hall voltage
measurement.
Data type is
NRf
<HallVoltageAverageInVolts> The average Hall voltage measured, in volts.
Data type is
NRf
<HallVoltageStandardErrorInVolts>
The standard error of the Hall voltage, in volts.
Data type is
NRf
<hallVoltageSnr>
The signal-to-noise ratio of the Hall voltage.
measurement.
Data type is
NRf
<carrierType>
The carrier type of the sample. P, N, or Unknown.
Data type is
NAMED
<pTypeCount>
The number of samples that were measured as P carrier
type.
Data type is
NR1
<nTypeCount>
The number of samples that were measured as N carrier
type.
Data type is
NR1
<CarrierConcentrationAveragePerMetersCubed>
The average bulk carrier concentration measured in units of
per meters cubed.
Data type is
NRf
<SheetCarrierConcentrationAveragePerMetersSquared>
The average sheet carrier concentration measured in units
of per meters squared.
Data type is
NRf
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...