![Lake Shore Cryotronics Measure Ready M91 FastHall User Manual Download Page 153](http://html.mh-extra.com/html/lake-shore-cryotronics/measure-ready-m91-fasthall/measure-ready-m91-fasthall_user-manual_670649153.webp)
143
|
www.lakeshore.com
Remarks
Queries the error/event queue for all the unread items and removes them from the
queue. If there are multiple errors queued, the response returns a comma separated
list of code, error message pairs, in FIFO order. If the queue is empty, the response
code is 0 and the error message is “No error”.
:CLEar
Command Format
SYSTem:ERRor:CLEar
Parameter
None
Examples
Command
SYST:ERR:CLE
Remarks
This command removes all errors from the error/event queue.
:COUNt?
Query Format
SYSTem:ERRor:COUNt?
Return Parameter
<count>
Number of errors in the queue
Data type is NR1
Examples
Query
SYST:ERR:COUN?
Query Response
4
Indicates there are 4 unread errors in the queue
Remarks
Returns the number of unread items in the error/event queue for the number of
unread items. If the queue is empty, the response is 0.
[:NEXT]?
Query Format
SYSTem:ERRor[:NEXT]?
Return Parameter
<code>,<error message>
code is the error event code
Data type is NR1
error message is a description of the error
Data type is string
Examples
Query
SYST:ERR?
Query Response
-113,”Undefined header;qwerty”
Remarks
Queries error/event queue for the next item and removes it from queue.
SYSTem:FACTORYRESET
Format
SYSTem:FACTORYRESET
Examples
SYST:FACTORYRESET
Remarks
This command resets the instrument settings to its factory defaults.
SYSTem:KLOCk
Command Format
SYSTem:KLOCk <on/off>
Parameter
<on/off>
0 = unlocked, 1 = locked
Data type is bool
Query Format
SYSTem:KLOCk?
Return Parameter
<on/off>
Examples
Command
SYST:KLOC 1
Locks the front panel from being used
Query
SYST:KLOC?
Query Response
0
Indicates that the front panel is unlocked
Remarks
This command locks the front panel controls of the instrument.
Summary of Contents for Measure Ready M91 FastHall
Page 6: ...MeasureReady M91 FastHall Measurement Controller ...
Page 10: ...MeasureReady M91 FastHall Measurement Controller ...
Page 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Page 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Page 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Page 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...