34
8: Configuration
This section provides instructions to access and understand the current configuration of
the biosafety cabinet, and make changes to the configuration.
Keypad button presses are shown as
[BLUE WITH BRACKETS]
. Menu screen
selections are shown as
green italics
.
Current Configuration
It is important to understand the current configuration of the biosafety cabinet for many
reasons, some of which include:
Verify model and type
Verify accessories installed
Troubleshooting
The Info Screen provides information on the current configuration of the biosafety
cabinets, and details the current blower speed(s). The Info Screen is not password
protected. To access this screen, follow these steps:
1. From the Main Menu, press
[MENU]
, using
[UP]
or
[DOWN]
select the
Tools
icon. Press
[OK/MUTE]
.
[MENU]
Tools