36
Change Configuration
Changes to the biosafety cabinet configuration can be made in the field. This typically
occurs after an accessory is field installed, or the user requests a change to the sash
height (between 8 and 10 inches) on a Type C1 only.
Never change the following configuration parameters:
Cabinet Type
– if the cabinet type is changed, it can result in incorrect operation
of the biosafety cabinet, and the biosafety cabinet may not protect the product,
personnel, or environment.
Sash Height
– On Type A2 and B2 cabinets, never change the sash height.
The following instructions detail all screens in the Configuration menu, which is
password protected. The password is
[LIGHT] [UV LIGHT] [TIMER] [TIMER]
[OK/MUTE]
.
1. From the Main Menu, press
[MENU]
, using
[UP]
or
[DOWN]
select the
Tools
icon. Press
[OK/MUTE]
.
2. Use
[UP]
or
[DOWN]
to select the
Password
icon, press
[OK/MUTE]
.
[MENU]
Tools
Password
[OK/MUTE]
WARNING !
RESTRICTED AREA
FOR TECHNICIAN
USE ONLY