128
10. The exhaust blower/motor assembly may be removed from the exhaust plenum
at this point by removing the several thread-cutting, hex head screws located on
the motor side of the exhaust plenum. Use a 5/16-inch socket or nut driver to
remove these screws. The blower/motor assembly must be rotated
approximately 45 degrees to orient it correctly for removal.
11. The blower/motor assembly must be rotated approximately 45 degrees to orient it
correctly for removal, as shown in Figure 12-41b.
Reinstallation Notes:
1. When upward pressure begins to be applied onto the exhaust HEPA, only tighten
until the filter is secure and the filter gasket has 50% compression. Do not
overtighten the bolts.
Figure 12-41a
Figure 12-41b
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...