157
15: Diagnostics
This section details diagnostic functions available through the software of the biosafety
cabinet.
Sash Sensors
1. From the Home Screen, press
[MENU]
.
2. Press
[UP]
to select the
Tools
option.
3. Press
[OK/MUTE]
to enter the submenu.
4. Press
[DOWN]
until the
Diagnostics
option is
highlighted, then
[OK/MUTE]
to enter the submenu.
5. The
Sash Sensors
option is the first option, press
[OK/MUTE]
to enter Sash Sensor diagnostics.
Note:
If the blower is on when entering the Sash
Sensor Diagnostic screen, it will automatically
turn off.
6. Slowly raise the sash from fully closed to open. When
closed, the screen will report
CLOSED
. Below or
above the proper sash height, the screen will report
NOT AT WORKING HEIGHT
. At the proper sash
height (+/- 0.75 inches), the screen will report
AT
WORKING HEIGHT
. Type A2 12-inch and C1 will also
report
ALMOST CLOSED
when sash 2-3 inches (5-8
cm) above closed.
7. Press
[MENU]
to return to the Diagnostic submenu.