75
2. Take the downflow velocity reading at each test point location of the test grid.
3. Average all individual test point velocity readings.
4. If the average is not within the acceptance criteria range, adjust the biosafety
cabinet according to
Adjusting the Blower Speed
and/or
Adjusting the Internal
Damper
found in
Section 9: Calibration
.
Acceptance Criteria
Average downflow shall be as specified in Table 10-6
column “Downflow Average” +/-5
FPM (+/-0.03 m/s), and all individual test points /-16 FPM (+/-0.081 m/s) from
the average.
Type
A2 & B2
Cabinet
Width
Sash
Height
Downflow
Average
1
Test Grid
Distance
in inches (cm)
inches
FPM (m/s)
Total Test
Points
Rows x
Columns
From Back
& Sides
Between
Rows
Between
Columns
3-foot
8, 9
55 (.28)
21
3 x 7
6.0 (15)
5.75 (14.6)
4.08 (10.4)
10
55 (.28)
21
3 x 7
6.0 (15)
5.55 (14.1)
4.08 (10.4)
4-foot
8, 9
55 (.28)
24
3 x 8
6.0 (15)
5.75 (14.6)
5.21 (13.2)
10
55 (.28)
24
3 x 8
6.0 (15)
5.55 (14.1)
5.21 (13.2)
12
60 (.30)
24
3 x 8
6.0 (15)
5.44 (13.8)
5.21 (13.2)
5-foot
8, 9
55 (.28)
30
3 x 10
6.0 (15)
5.75 (14.6)
5.39 (13.7)
10
55 (.28)
30
3 x 10
6.0 (15)
5.55 (14.1)
5.39 (13.7)
12
60 (.30)
30
3 x 10
6.0 (15)
5.44 (13.8)
5.39 (13.7)
6-foot
8, 9
55 (.28)
36
3 x 12
6.0 (15)
5.75 (14.6)
5.50 (14.0)
10
55 (.28)
36
3 x 12
6.0 (15)
5.55 (14.1)
5.50 (14.0)
12
60 (.30)
36
3 x 12
6.0 (15)
5.44 (13.8)
5.50 (14.0)
Type C1
Cabinet
Width
Sash
Height
Downflow
Average
1
Test Grid
Distance
in inches (cm)
inches
FPM (m/s)
Total Test
Points
Rows x
Columns
From Back
& Sides
Between
Rows
Between
Columns
4-foot
8
65 (.33)
24
3 x 8
6.0 (15)
5.75 (14.6)
5.21 (13.2)
10
65 (.33)
24
3 x 8
6.0 (15)
5.55 (14.1)
5.21 (13.2)
6-foot
8
55 (.28)
36
3 x 12
6.0 (15)
5.75 (14.6)
5.50 (14.0)
10
55 (.28)
36
3 x 12
6.0 (15)
5.55 (14.1)
5.50 (14.0)
1. Tolerance for Average Downflow Velocity = +/- 5 FPM (+/- 0.03 m/s)
Table 10-6
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...