Inflow Velocity
All models of Labconco biosafety cabinets are tested per the current NSF/ANSI
Standard 49. This section describes how to prepare and test the inflow velocity on all
Types of biosafety cabinets.
The front Dress Panel must be installed on the biosafety cabinet to obtain accurate
inflow results. If the biosafety cabinet is connected to a remote (building) exhaust
system, ensure the remote exhaust system is on.
Primary Method
NSF/ANSI Standard 49 now requires the Primary Method for determining inflow be
utilized unless there is less than 41 inches (104 cm) clearance between the front of the
biosafety cabinet and the nearest physical, permanent obstruction.
The Primary Method requires the following:
DIM (Direct Inflow Meter), such as a Shortridge
TM
Labconco Holder Bracket Assembly (Catalog Number 3836405), shown below in
Figure 10-28.
Thin, rigid plastic (or similar material) blocking plates, to close off front work
opening on either side of the DIM so all inflow air is directed through the DIM.
Removable tape to secure the plastic blocking plates.
1. Start the blower(s), ensure the sash is at the correct operating height.
2. Attach the Labconco Holder Bracket Assembly to the front grille centered left-to-
right on the biosafety cabinet. Loosen the two Thumbnuts to create a gap
between the two plates.
3.
Place the DIM’s skirt frame into the gap between the two plates on the Bracket
Assembly.
Figure 10-28
Two pins insert into front-most
row of holes on the front grille
Thumbnuts
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...