10. Grasp the motor and pull the assembly straight out of the biosafety cabinet. See
Figure 12-31.
Reinstallation Notes:
1. During reinstallation of the blower/motor assembly, angle the rear blower bracket
into the plenum opening. Make sure this rear blower bracket is INSIDE the
plenum opening. Once reinstalled, and the three blower mounting bolts are
installed and tightened, verify the rear blower bracket is positioned correctly by
trying to rotate/move the blower/motor assembly. If the entire assembly is rigid
(does not move in any direction), the rear blower bracket is properly secured.
Figure 12-31