170
8. This screen allows you to set the Type of cabinet.
Leave this setting alone, and press
[OK/MUTE]
.
9. This screen allows you to configure the cabinet for an
airflow sensor or not. If you are not changing the
selection, press
[OK/MUTE]
.
10. This screen configures the unit for a UV light or not.
Leave this setting alone, and press
[OK/MUTE]
.
11. This screen allows you to configure the cabinet for a
canopy or not. Use
[UP]
or
[DOWN]
to select
Canopy
.
After your selection is chosen, press
[OK/MUTE]
.
12.
This screen allows you to configure the cabinet’s sash
height. Leave this setting alone, and press
[OK/MUTE]
.
The unit will then return to the Home Screen in the new
configuration.
13. Take Inflow and Downflow readings as described in
Section 9: Calibration
. If not
within acceptable tolerance for the model, adjust the biosafety cabinet’s internal
damper and/or blower(s) speed until Inflow and Downflow are correct. Use the
original blower speed(s) as a reference.