Airflow Sensor (optional)
1. From the Home Screen, press
[MENU]
.
2. Press
[UP]
to select the
Tools
option.
3. Press
[OK/MUTE]
to enter the submenu.
4. Press
[DOWN]
until the
Diagnostics
option is
highlighted, then
[OK/MUTE]
to enter the
submenu.
5. Press
[DOWN]
until the
Airflow Sensor
option is
highlighted, then
[OK/MUTE]
to enter Airflow
Sensor diagnostics.
6. The display shows the status of the airflow
sensor in the following order:
current sensor output in millivolts,
current inflow, corrected for the display
zero set-point in millivolts
alarm set-point in millivolts
nominal (105 FPM or 0.53 m/s) output in
millivolts.
7. Press
[MENU]
to return to the Diagnostic submenu.