15. Average all the readings. Apply the wind tunnel correction and local condition
corrections factors to the average velocity reading.
16. Multiply the average inflow velocity by the column
“Correction Factor” based on
Type and model in Table 10-8.
Type
A2 & B2
Cabinet
Width
Sash
Height
Sash
Stop
Template
Probe
Distance
# of Test
Points
Correction
Factor
Corrected Inflow
Volume Range
inches
inches
(mm)
CFM
m
3
/hr
3-foot
8
3+
4 (102)
6
0.95
203-223
345-379
9
3+
4 (102)
6
1
228-251
387-426
10
3+
4 (102)
6
0.95
253-278
430-472
4-foot
8
4+
3.25 (83)
8
1
269-296
457-503
9
4+
3.25 (83)
8
1
303-333
515-566
10
4+
3.25 (83)
8
1
337-371
573-630
12
4+
3.25 (83)
8
1.08
404-444
686-754
5-foot
8
5+
3.25 (83)
10
1
336-370
571-629
9
5+
3.25 (83)
10
1
378-416
642-707
10
5+
3.25 (83)
10
1
420-462
714-785
12
5+
3.25 (83)
10
1
505-555
858-943
6-foot
8
6+
4 (102)
12
1.83
403-443
685-753
9
6+
4 (102)
12
1.83
453-498
770-846
10
6+
4 (102)
12
1.83
503-553
855-940
12
6+
4 (102)
12
2.0
606-666
1030-1132
Type C1
Cabinet
Width
Sash
Height
Sash
Stop
Template
Probe
Distance
# of Test
Points
Correction
Factor
Corrected Inflow
Volume Range
inches
inches
(mm)
CFM
inches
4-foot
8
4+
3.25 (83)
8
1.08
269-296
457-503
10
4+
3.25 (83)
8
1.08
337-371
573-630
6-foot
8
6+
4 (102)
12
2
403-443
685-753
10
6+
4 (102)
12
2
503-553
855-940
Table 10-8
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...