IM-A-1048H
51
No.
Functions
Initialize
Parameters
Digits
Options/ranges
Initial
values
Remarks
F-16
RS
communication
settings
For details, see the
"INSTRUCTION MANUAL (FOR
BCD/RS-232C/RS-485/TEDS
OPTIONS)".
F-17
Device ID
For details, see the
"INSTRUCTION MANUAL (FOR
BCD/RS-232C/RS-485/TEDS
OPTIONS)".
F-18
CC-Link
For details, see the
"INSTRUCTION MANUAL (FOR
CC-LINK OPTION)".
F-19
Numeric value
registering
calibration
Zero-point
output value
-3.2000 to +3.2000
0.0000
The ZERO indicated value that is
registered at calibration.
F-20
Operation
pattern
Operation
pattern
10e0
0 Pattern 0
0
Selects the pattern of comparator
values from registered 4 patterns.
1 Pattern 1
2 Pattern 2
3 Pattern 3
F-21
Updating rate of
displaying
values
Updating
rate
10e0
0 15.60 times/sec
0
Thinning: 1/256 time, 64msec
1
7.80 times/sec
Thinning: 1/512 time, 128msec
2
3.90 times/sec
Thinning: 1/1024 time, 256msec
3
1.95 times/sec
Thinning: 1/2048 time, 512msec
4
0.98 times/sec
Thinning: 1/4096 time, 1024msec
5
0.49 times/sec
Thinning: 1/8192 time, 2048msec
6
0.24 times/sec
Thinning: 1/16384 time, 4096msec
7
0.12 times/sec
Thinning: 1/32768 time, 8192msec
F-22
Level Test
Signals
Display
additional
function
10e0
0 Disabled
0
Outputs the LEVEL TEST values
only.
1 Enable
Outputs a value (current indicated
value + LEVEL TEST value).
Level Test
values
-99999 to +99999
0
Level Test
ON/OFF
10e0
0 OFF
0
Turns OFF the LEVEL TEST
values.
1 ON
Turns ON the LEVEL TEST values.
1 Low limit
F-23
[HH]
comparator
(comparator-3)
functions
"High limit"
or "Low
limit"
10e0
0 High limit
0
Assign "0: High limit" or "1: Low
limit" of the [HH] comparator
(comparator-3).
1 Low limit
F-24
[HI] comparator
(comparator-1)
functions
"High limit"
or "Low
limit"
10e0
0 High limit
0
Assign "0: High limit" or "1: Low
limit" of the [HI] comparator
(comparator-1).
1 Low limit
F-25
[LO]
comparator
(comparator-2)
functions
"High limit"
or "Low
limit"
10e0
0 High limit
1
Assign "0: High limit" or "1: Low
limit" of the [LO] comparator
(comparator-2).
1 Low limit
F-26
[LL] comparator
(comparator-4)
functions
"High limit"
or "Low
limit"
10e0
0 High limit
1
Assign "0: High limit" or "1: Low
limit" of the [LL] comparator
(comparator-4).
1 Low limit
F-27
Display
stability
Disable/
Enable
10e0
0 Disable
0
1 Enable